How to Restart an HTEC System After a Long-Term Shutdown
Purpose
The procedure outlined here would be applicable for an HTEC that has been out of use for over several weeks.
Solution
Before restarting an HTEC, please inspect to ensure proper shut down procedures were performed. Most issues dealing with improper storage are related to salt precipitates clogging lines when mobile phase dries out, or bacteria growth, For proper shutdown of the HTEC-500/510, please see this article. Signs of leaks or clogging include extremely high or low pressure. A spike in pressure reading followed by a rapid drop to zero indicates a clog followed by a fitting becoming loose.
Please watch these videos we have prepared for HTEC Restart Procedures:
Part 1: Plumbing
Part 2: ECD Cell
Part 3: Fittings